| |
|
Amiga Video Toaster
/ Flyer FAQs
Amiga Flyer
Calibration Topics
Manual Hue Phase Adjustment Procedures for
Flyers
Product: Flyer
Platforms: 2000 3000 3000T 4000 4000T
These programs require the following hardware:
-
A vectorscope on internal sync at 75% bars connected
to the flyer MAIN OUTPUT BNC with 75 ohm termination.
-
An NTSC Signal Generator on 75% SMPTE bars connected
to the Toaster's VID 1 BNC input. Input termination
may be verified by pressing the F10 function key
to see the check at the input 1 box.
-
An AUTOHUED Toaster.
-
A Flyer with video drives on SCSI0 (FlyerA or
FA0) and SCSI1 (FlyerB or FB0).
-
An Amiga 2000 or 4000 loaded with 4.1 Flyer software.
These programs require the following clip and framestore:
-
The "107.FS.IVectorPattern" framestore
which comes with the ManPhase.lha distribution
should be put into the NewTek/Framestore drawer
by the install procedure which is part of the distribution.
>> Load the framestore 107.FS.IVectorPattern
into DV1 of the Toaster.<<
-
The "wd2-bars" video clip must be recorded
onto SCSI0 (FlyerA or FA0) and SCSI1 (FlyerB or
FB0) of the Flyer system.
HUEPA and HUEPB
These programs perform manual Play Hue Phase adjustment,
with HUEPA for channel A and HUEPB for channel B.
-
Load the Toaster-Flyer software and do a FlyerCalibrate
so the Flyer will pick an aligner edge.
-
Double-click on the Workbench icon for HUEPA.
A window appears with 4 sliders.
-
The top slider, POSITION, should be set to -35.
-
Put the remaining DEGREE sliders to the far left.
-
Move the CLK(22.5 degree) slider to the right
until the vectorscope display's color points are
at the nearest counter-clockwise (CCW) position
from the center of the small boxes on the vectorscope.
Red and magenta would be left of the small boxes.
If it is possible to set the color points in the
exact center of the small boxes do so and go to
step 8. If when setting the CLK(22.5 degree) slider,
the vector display is at or past 90 degrees CW
(clockwise), the next click may make it jump past
the small boxes. In this case, go to step 6 now.
-
Move the COARSE(4 degree) slider to the right
until the vectorscope display's color points are
at the nearest counter-clockwise (CCW) position
from the center of the small boxes on the vectorscope.
Red and magenta would be left of the small boxes.
If it is possible to set the color points in the
exact center of the small boxes, do so.
-
Move the FINE(1 degree) slider to the right until
the vector display is nearest to the center of
the small boxes.
-
Check the phase linearity by moving the COARSE(4
degree) slider left and right 2 clicks each way.
If the vector pattern jumps more than the selected
amount (like 45, 60, or 90 degrees) start the procedure
over and during step 2 try using the COARSE (4
degree) slider one click to the right while adjusting
the CLK(22.5 degree) slider. As in step 7, use
the FINE(1 degree) to make the final adjustment.
-
In cases which do not adjust with linearity, rerun
FlyerCalibrate to reselect the aligner edge. If
this fails to obtain phase linearity as the COARSE(4
degree) slider is moved back and forth 2 clicks
(as described in step 8) it may be necessary to
accept a linearity of one click each way. If phase
linearity problems persist see the note at the
end of this file. You may need to contact NewTek
Tech support for a hardware repair.
-
Repeat HUEPA quickly as settings may drift as
board temperature stablizes. Channel B may look
like it has proper color phase during the clip
playback section of HUEPA, but it tends to track
channel A unless it is way off. Be sure to run
HUEPB to check and set channel B as shown in step
11.
-
Perform steps 2 through 10 using the HUEPB program
to set video channel B's play color phase.
-
If not going on to HUER for record phase adjustment,
and if needing to turn the system off be sure to
exit the Flyer software from the Editor/Switcher
using SHUTDOWN in the VIEWS menu to prevent loss
of settings.
HUER
The record hue phase adjustment is set with HUER by
double-clicking on the HUER icon. A window appears
with 4 sliders.
-
The 107.FS.IVectorPattern framestore should appear
on the video monitor. It has a large orange stripe
down the middle of the monitor. If it does not
appear, load it into DV1 from the NewTek/Framestore
drawer. You may need to use Left-AMIGA M to switch
screens to get to the Switcher/Editor and back
to HUER.
-
The top slider is POSITION and should be set to
22 on a good Flyer.
-
Put the Clk(22.5 degree) slider fully left, the
Coarse(4 degree) slider fully right and the Fine(1
degree) slider fully left.
(Note: This is different from
HUEPA and HUEPB).
-
The Clk(22.5 degree) slider has 4 positions. The
object is to try each one to find a "sweet
spot" by moving the Coarse(4 degree) slider
from right to left to locate a linear phase region
around 0 degrees as displayed near the bottom of
the Record Phase Test window. The best regions
will be linear for plus and minus at least 8 to
10 degrees either side of zero. The Coarse(4 degree)
slider should change the displayed degrees by roughly
4 degrees for each position.
-
If the displayed degrees jump well past the expected
4 degrees (like 45, 60, or 90 degrees), you must
try the next Clk(22.5 degree) slider position to
the right and repeat step 4. Try each of the Clk(22.5
degree) positions until a stable region is located
by scanning it with the Coarse(4 degree) slider.
-
If a region is found to be almost but not quite
linear for 8 to 10 degrees either side of zero,
the Fine(1 degree) slider may be used to shift
the region slightly to one side or the other.
-
Finally, the Fine(1 degree) slider is used to
bring the displayed degrees nearest to zero.
-
After using step 7, recheck the region's phase
linearity by scanning it with the Coarse(4 degree)
slider to see that it is stable up and down about
8 to 10 degrees, and leave it set nearest to 0
degrees. If the displayed phase jumps about 45,
or 60 degrees try other positions of the 1 degree
slider and then reset the Coarse(4 degree) slider,
and scan up and down about 8 to 10 degrees. This
can affect the linearity of the region around 0
degrees.
-
If still no linear region is found, rerun FlyerCalibrate
and try again. This will also strip the HUEPA and
HUEPB settings, so rerun them before doing HUER.
If no linear region is found after about 4 FlyerCalibrates
and HUERs, try the note below. You may need to
contact the NewTek Tech department for hardware
repair.
-
Click the close gadget in the top left of the
Record Phase Test window to check the results.
Although channel A and channel B playing wd2.bars
should play right in the small vectorscope boxes
if properly adjusted, the third clip of all three
color phase programs will only look as phase accurate
as the incoming signal from your signal generator.
Any observed shift in the incoming signal during
recording should be matched during the playback
of the third clip which is the RECORDED clip for
HUER. (You can distinguish the incoming pattern
out of the signal generator by the STRAIGHT lines
connecting the color dots. Digital playback of
color bars shows a bending or skewing between color
dots during retrace. This is normal and does not
indicate a Flyer problem).
-
To shut off the AMIGA, exit from the Editor/Switcher
using SHUTDOWN in the VIEWS menu of the Editor
to save your settings.
Note: Video cable lengths and terminating
impedances DO affect phase adjustment in video systems.
It is possible to change the results of the above adjustments
by changing video cable lengths or terminating video
devices on the main video output. You may verify your
terminating impedance by measuring it directly with
a digital ohmmeter at the BNC cable connected to the
Flyer Main Output. Disconnect the cable from the Flyer
Main Output. With the meter set to Ohms, touch the
black lead to the shell of the cable’s BNC connector
and touch the red lead to the pin in the center of
the cable’s BNC connector. The reading should
be VERY close to 75 ohms.
|
|
|
|
|
|